Z-1214-2018: Sonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560

Recall record Z-1214-2018 for Konica Minolta Medical Imaging USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1214-2018
Company in FDA record
Konica Minolta Medical Imaging USA, Inc.
FDA record creation date
Mar 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.

Source facts

Action
Konica Minolta is preparing for the final AC Adapter replacement. Upon receipt of the new AC Adapter, owners of SONIMAGE HSl System will be asked to replace the existing AC Adapter. Additionally, instructions for the return of the current AC Adapter will be provided.
Initiated
2017-12-04
Root cause
Device Design
Terminated
2019-10-28
Distribution
Distributed in the following U.S. states: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI and the District of Columbia. Distributed in Canada and Brazil.
Recall number
Z-1214-2018
Recall status
Terminated
CDRH record ID
162379
FDA record created
2018-03-27
Recall event number
79455

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗