Z-1214-2023: epoc Host 2 as follows: Product SMN epoc Host 10736387 epoc Host 2, Canada 10736388 epo

Recall record Z-1214-2023 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1214-2023
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA classification date
Mar 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

Source facts

Action
Siemens has issued an Urgent Medical Device Correction POC 23-004.A.US on December 12, 2022 and Urgent Field Safety Notice POC 23-004.A.OUS to customers informing them of the issue and providing workaround for customers to use until the affected sensor configuration/software version expires on June 12, 2023. The next sensor configuration with the fix and its corresponding sw version will be available to customers in April 2023. Letter states reason for recall, health risk and action to take: " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " If you are a distributor, please ensure your customers receive this UMDC letter. " If you perform QC tests at barometric pressures equal to or above 730 mmHg (97 kPa), no additional action is required. " If you perform QC tests at barometric pressures less than 730 mmHg (97 kPa), apply barometric correction factors before comparing the gas readings with the published value assignments. ***Important Note*** The above barometric correction factor should be applied only to epoc System Software sensor configuration 41.1, epoc Host 2 SW Version 3.37.3 and epoc NXS Host SW Version 4.10.6, which expires June 12, 2023. Following the expiration of this sensor configuration and the software version refer to the epoc System Manual Section 9.5.1 for the appropriate Barometric Correction factor to apply." Please review this letter with your Medical Director." Please retain this letter with your laboratory records and forward this letter to those who may have received the affected product. Questions, please contact your Siemens Healthineers Technical Solution Center or your local Siemens Healthineers technical support representative.
Initiated
2022-12-12
Root cause
Under Investigation by firm
Distribution
Nationwide Foreign: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kuwait Laos Latvia Lithuania Luxembourg Malaysia Mali Martinique Mexico Myanmar Netherlands New Caledonia New Zealand Norway Papua Nw Guinea Peru Philippines Poland Portugal Qatar Republic Korea Reunion Romania Russian Fed. San Marino Singapore Slovakia Slovenia Solomon Islands South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey U.A.E. Ukraine United Kingdom Uruguay Vietnam
Recall number
Z-1214-2023
Recall status
Open, Classified
CDRH record ID
198394
FDA classified
2023-03-03
Recall event number
91639

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗