Skip to content

Public FDA record

Z-1215-2015: The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-g

Recall record Z-1215-2015 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1215-2015
Company in FDA record
Brainlab AG
FDA classification date
Feb 26, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.

Source facts

Action
FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated December 11, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to utilize the sterilization cycles included in the letter. Customers were further advised to alert users of these devices to the sterilization process changes. Customers with questions can contact Brainlab by telephone at 800-597-5911 or via e-mail at us.support@brainlab.com.
Initiated
2015-01-16
Root cause
Reprocessing Controls
Terminated
2017-03-13
Distribution
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
Recall number
Z-1215-2015
Recall status
Terminated
CDRH record ID
132865
FDA classified
2015-02-26
Recall event number
70291

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗