Open original FDA record ↗Recall record
- FDA identifier
- Z-1216-2007
- Company in FDA record
- Summit Industries Inc.
- FDA classification date
- Aug 29, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The D800 Swivel Collimator may come loose from the tubestand and fall onto the x-ray table.
Source facts
- Action
- The affected dealers were notified of the field correction recall by letter dated 8/3/07, informing them of the problem with the stripped screws and requesting that they notify their affected end user locations. The end user notification letters inform them of the problem and state that the dealer will be contacting them soon to arrange a time to visit the end user location and conduct the rework on the collimator, which should take less than 30 minutes. In the mean time the end user was instructed to examine their swivel collimator for looseness. If it is found to be loose, the end user was instructed to take the unit out of service and call the dealer.
- Initiated
- 2007-08-10
- Root cause
- Other
- Terminated
- 2010-03-18
- Distribution
- Worldwide, including USA, Australia, Cambodia, Canada, Chile, Colombia, Dominican Republic, Ecuador, France, Germany, Haiti, Mexico, Nicaragua, Singapore, and South Africa.
- Recall number
- Z-1216-2007
- Recall status
- Terminated
- CDRH record ID
- 53880
- FDA classified
- 2007-08-29
- Recall event number
- 38512