Z-1216-2010: AccuMax Quantum Complete (Model CU3/CU4), an alternating pressure with turn assist & on demand low air lo
Recall record Z-1216-2010 for Encompass Therapeutic Support Systems dba BG North America. View source facts, labeled dates, and the original FDA record.
Encompass Therapeutic Support Systems dba BG North America
FDA classification date
Mar 31, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Potential out gasing of the battery and short term ignition of the gas within the control unit enclosure. This failure mode includes a risk for fire.
Source facts
Action
An "URGENT RECALL NOTIFICATION" letter dated March 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers should immediately discontinue use of the control unit, turn off the unit and disconnect from the wall socket and quarantine the unit away from any patient's room. The customers should immediately examine your inventory and quarantine the products listed; identify and notify any facility where they may have further distributed or transferred the listed products; and complete and return the REPLY/FAX-BACK FORM as soon as possible to Fax: (626) 796-7856 or mail to EncompassTSS, 100 E. Corson Street, Ste 310, Pasadena, CA 91103.
Please contact EncompassTSS Customer Service at 800-822-8288 if you have any questions regarding this recall notice.
Initiated
2010-03-05
Root cause
Nonconforming Material/Component
Terminated
2012-03-21
Distribution
Nationwide distribution: CA, TX, IN, OR, and MO.
Recall number
Z-1216-2010
Recall status
Terminated
CDRH record ID
89827
FDA classified
2010-03-31
Recall event number
55074
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.