Stryker Instruments and Stryker Leibinger GmbH & Co. KG are
voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker Navigation
System because the settings on
the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle
Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.
Source facts
Action
US customers will be notified via FedEx overnight. US Sales Representatives will be notified via email. International entities will be notified via an internal notification system. International customers will be notified per local regulations/procedures.
Customer Instructions:
Actions to be taken by the Customer/User:
1. Immediately review this Recall Notification.
2. Immediately check all stock areas and/or operating room storage for affected equipment. Quarantine and
discontinue use of any affected Thoracic Pedicle Feelers.
3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this Notification and identify how many
affected items are currently in your inventory. Please complete and return the BRF even if you don't have any affected product on hand.
Note: Your signature on the BRF indicates that you received and understand this Notification and have followed the
instructions in the Notification.
4. If you have further distributed this product, please forward this Notification and the attached BRF to all affected locations. Please indicate each location on the BRF.
5. If the BRF for your facility indicates that recalled product is currently on hand, we will contact you to arrange for the return of the product for repair/replacement.
If discontinuing use of the specific Thoracic Pedicle Feeler(s) listed will cause disruption at your facility. For further questions, please call (269) 323-7700.
Initiated
2017-10-16
Root cause
Software Manufacturing/Software Deployment
Terminated
2019-04-03
Distribution
International , US and US Govt.
Recall number
Z-1216-2018
Recall status
Terminated
CDRH record ID
162049
FDA record created
2018-03-27
Recall event number
79335