If cables are improperly connected, current may reach high charge density, causing tissue harm.
Source facts
Action
The firm originally notified customers of the issue via email on September 4, 2018.
Also, another updated notification Urgent Field Safety Notice, Neuro Omega system, DATE March 4, 2019, Field Advisory Note [FAN]
Attention: Details on affected devices: Neuro Omega system with: Catalogue number: (Neuro Omega) NOM-000000-00 + (Drive) 750-000020-00 + (EMG) 750-000010-00 + (EEG) 750-000010-12 Software version number: Any,
clarifying that the issue only occurred in a triple-fault condition.
Initiated
2018-09-04
Root cause
Device Design
Terminated
2020-11-03
Distribution
US nationwide distribution: AL, AZ, CA, CO, GA, IL, LA, MD, MA, MI, MN, OH, PA, RI, TX, and VA
Recall number
Z-1216-2019
Recall status
Terminated
CDRH record ID
171800
FDA classified
2019-05-06
Recall event number
82540