Skip to content

Public FDA record

Z-1216-2022: Ratcheting Screwdriver Handle, provided in the Aequalis Perform Reversed kit

Recall record Z-1216-2022 for Tornier, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1216-2022
Company in FDA record
Tornier, Inc
FDA record creation date
Jun 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device is unable to meet sterilization requirements when sterilized per instructions in an Aesculap container.

Source facts

Action
The recalling firm issued recall notification letters on May 3, 2022 via UPS ground. Consignees are asked to do the following: 1. Immediately cease sterilizing the kitted Aequalis" Perform" Reversed Ratcheting Screwdriver Handle (MWJ128) in an Aesculap container. a. If sterilization in an Aesculap container is desired, the handle (MWJ128) shall be removed from the kit (YKAD261) prior to sterilization and then sterilized separately. To ensure sterility, the handle (MWJ128) must be double-wrapped and sterilized individually according to the parameters provided in the table below and provided within the new Instructions for Use online at https://ifu.stryker.com/. b. If the kit, YKAD261, contains the previous version (IFU 7638), discard the previous IFU and reference the eIFU (IFU 017) for instructions. 2. Disseminate this notice to all those who need to be aware within your organization. 3. Return the enclosed business reply form by email to confirm receipt of this notification. 4. Inform Stryker if any of the subject devices have been distributed to other organizations.
Initiated
2022-05-03
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-1216-2022
Recall status
Open, Classified
CDRH record ID
193256
FDA record created
2022-06-01
Recall event number
90052

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗