Open original FDA record ↗Recall record
- FDA identifier
- Z-1216-2022
- Company in FDA record
- Tornier, Inc
- FDA record creation date
- Jun 1, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The device is unable to meet sterilization requirements when sterilized per instructions in an Aesculap container.
Source facts
- Action
- The recalling firm issued recall notification letters on May 3, 2022 via UPS ground. Consignees are asked to do the following: 1. Immediately cease sterilizing the kitted Aequalis" Perform" Reversed Ratcheting Screwdriver Handle (MWJ128) in an Aesculap container. a. If sterilization in an Aesculap container is desired, the handle (MWJ128) shall be removed from the kit (YKAD261) prior to sterilization and then sterilized separately. To ensure sterility, the handle (MWJ128) must be double-wrapped and sterilized individually according to the parameters provided in the table below and provided within the new Instructions for Use online at https://ifu.stryker.com/. b. If the kit, YKAD261, contains the previous version (IFU 7638), discard the previous IFU and reference the eIFU (IFU 017) for instructions. 2. Disseminate this notice to all those who need to be aware within your organization. 3. Return the enclosed business reply form by email to confirm receipt of this notification. 4. Inform Stryker if any of the subject devices have been distributed to other organizations.
- Initiated
- 2022-05-03
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1216-2022
- Recall status
- Open, Classified
- CDRH record ID
- 193256
- FDA record created
- 2022-06-01
- Recall event number
- 90052