Skip to content

Public FDA record

Z-1217-2014: Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic

Recall record Z-1217-2014 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1217-2014
Company in FDA record
Exactech, Inc.
FDA classification date
Mar 18, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.

Source facts

Action
Exactech sent an "URGENT PRODUCT RECALL NOTICE" dated January 6, 2014, to all affected customers. The letter identified the product and action needed to taken by the customers. Customers were instructed to: 1) Immediately cease distribution. 2) Extend the information to accounts in possession of the product. 3) Verify if they have any of the subject Gibralt Screws from the specified lots. 4) Complete and fax back the attached form. 5) Return affected products. For further questions, please call ( 352) 327-4847.
Initiated
2014-01-06
Root cause
Mixed-up of materials/components
Terminated
2014-08-04
Distribution
USA Distribution including the states of FL, NY, OR, IL, PA, MO and NJ.
Recall number
Z-1217-2014
Recall status
Terminated
CDRH record ID
125540
FDA classified
2014-03-18
Recall event number
67486

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗