Z-1217-2026: BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Su

Recall record Z-1217-2026 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1217-2026
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Jan 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network

Source facts

Action
On 11/06/2025, the firm sent via mail and email an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter informing customers that during an electronic medical record interoperability workflow, a BD Alaris" PCU (Point of Care Unit) connected to an affected BD Alaris" Pump Module may disconnect from the hospital network. The connectivity issue relates to an unrecognized date and time ( datetime ) stored within the affected BD Alaris" Pump Module software. The software that was loaded onto the affected pump modules has a datetime that falls within the Daylight Savings Time (DST) adjustment. The Alaris" System Manager software that mediates connectivity between the PCU module and the hospital network does not recognize this DST datetime as valid, resulting in loss of connectivity between the PCU module and the network. Instructions to Clinicians: 1. If an affected Pump Module is being used in bi-directional interoperative workflows, return it to Biomedical Engineering when clinically feasible. 2. No immediate action is required if the facility does not utilize bi-directional interoperative workflows. In such cases, the affected Pump Module should be returned to Biomedical Engineering when not in use. Instructions for Customers and Distributors: 1. Locate the affected devices identified in Attachment A and return to Biomedical Engineering for re-flashing. 2. Complete the attached Customer Response Form and return it to the BD contact noted on the form so that BD may acknowledge your receipt of this notification. 3. Please contact the BD Recall Support Center to coordinate self-remediation or depot-based remediation. If performing self-remediation, the Support Center will provide a tracking form to document completion back to BD. Alternatively, customers may return their affected devices to be serviced by the BD Service Center. For Questions - contact Recall Support Center Phone: 1-888-562-6018 Phone Hours: 7:00am to 4:00pm PT, Monday- Friday Email: SupportCenter@bd.com
Initiated
2025-11-06
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of AZ, IN, OH, and VA.
Recall number
Z-1217-2026
Recall status
Open, Classified
CDRH record ID
216847
FDA classified
2026-01-29
Recall event number
97871

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗