Open original FDA record ↗Recall record
- FDA identifier
- Z-1218-06
- Company in FDA record
- Boston Scientific
- FDA classification date
- Jul 12, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
Source facts
- Action
- On May 30, 2006, Boston Scientific notified customers of the IQ Guide Wire manufactured from February 3, 2006, through March 15, 2006. BSC requested a reply whether or not they still have product in their institution. Replacements will be issued for returned product. UPDATE: A second recall notification was sent to affected customers 07/07/06. This second notification included Inner Pouch Number (UPN) and Inner Pouch Lot/Batch #.
- Initiated
- 2006-05-30
- Root cause
- Other
- Terminated
- 2007-06-28
- Distribution
- Nationwide (AR, AZ, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, NC, NH, NJ, NM, NV, OH, OK, PA, TN, TX, UT, VA, WI, WV) and Puerto Rico. Chile, Germany, Taiwan.
- Recall number
- Z-1218-06
- Recall status
- Terminated
- CDRH record ID
- 46387
- FDA classified
- 2006-07-12
- Recall event number
- 35549