Z-1218-2018: IMRIS ORT300 Removable Operating Room Table, 50-60 Hz Version is intended for use during diagnostic exami

Recall record Z-1218-2018 for Deerfield Imaging, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1218-2018
Company in FDA record
Deerfield Imaging, Inc.
FDA record creation date
Mar 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The rotational locking mechanisms on the ORT 200/300 tables fail to operate such that it cannot change states between locked and unlocked.

Source facts

Action
The firm disseminated their notices on 02/23/2018 by mail and followed with an email on 02/26/2018. The notice stated: "It has recently come to our attention that the ORT200/300 rotational lock fails to operate such that it cannot change states between locked and unlocked. This may pose a risk of injury to the patient. To minimize the risk of injury, IMRJS is taking the following corrective actions: " ORT200 and 300 tables: all customer sites with these tables shall have the rotational lock operation verified and adjusted, if necessary. IMRIS has identified that your facility has one or more of these OR Tables which require evaluation. Please confirm details on page 2 and return your response to IMRIS. IMRIS Customer Service will contact you to implement this corrective action. IMRIS recommends these preventive actions: " Per the OAT Operator Manuals the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. " During usage, verify the table is locked by manually pushing the table to make sure it does not move rotationally. " If the rotational lock is observed to be not functioning properly, please contact IMRIS Customer Service. The appropriate competent authorities have been notified of this corrective action. IMRIS is committed to patient and user safety, and apologizes for any inconvenience that this may cause. Please contact customer support via our service line at 1-866-475-0525 or Email: customersupport@imris. com if you have any questions or concerns."
Initiated
2018-02-23
Root cause
Under Investigation by firm
Terminated
2019-08-27
Distribution
US, Canada, China, Sweden, Germany, Japan, Qatar, France, Audtralia
Recall number
Z-1218-2018
Recall status
Terminated
CDRH record ID
162137
FDA record created
2018-03-27
Recall event number
79366

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗