Open original FDA record ↗Recall record
- FDA identifier
- Z-1218-2019
- Company in FDA record
- Smiths Medical ASD Inc.
- FDA record creation date
- Apr 25, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Several lots of Jelco ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.
Source facts
- Action
- Smiths Medical delivered an "URGENT MEDICAL DEVICE RECALL NOTICE" dated 8/22/2018 and accompanying Recall Response Forms via tracked mail on 08/22/2018 to all consignees who purchased potentially affected product. The notices describe the product, problem and actions to be taken. The consignees were instructed to do the following: 1. Locate affected Jelco ViaValve" safety I.V. catheters in your possession 2. Determine the number of affected devices in your possession and complete the Recall Response Form within 10 days of receipt, returning it to SmithsMedical6697@stericycle.com. The form must be returned even if you do not have any of the affected product in your possession. Product credit and/or replacement will be processed once the Recall Response Form is received and returned affected product has been verified. 3. All affected product must be returned to Stericycle for processing. Please include a copy of the completed Recall Response Form inside EACH BOX of returned product to facilitate processing. 4. DISTRIBUTORS: If you have distributed potentially affected product to your customers, please immediately notify them of this recall. If you have any questions, contact the Compliance Manager at 763-383-3310 or email to: fieldactions@smiths-medical.com.
- Initiated
- 2018-08-22
- Root cause
- Process design
- Terminated
- 2021-02-11
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, and New Zealand.
- Recall number
- Z-1218-2019
- Recall status
- Terminated
- CDRH record ID
- 166806
- FDA record created
- 2019-04-25
- Recall event number
- 80887