- FDA identifier
- Z-1221-2010
- Company in FDA record
- Leica Microsystems, Inc.
- FDA classification date
- May 6, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A manufacturing defect was identified that could potentially result in the laser shutter not closing as intended which could result in the shutter not completely closing disrupting the laser beam safely.
Source facts
- Action
- A Service Note was sent to service organizations on November 12, 2009. The communication asked customers to check if they must exchange their shutters and send their planned exchange date by 2009-11-20 to clsm.qc@leica-microsystems.com. Customers were also asked to check the serial numbers of the one-wire chips of the shutters. Then lastly, the were asked to exchange the shutters and confirm the exchange.
- Initiated
- 2009-11-12
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2010-05-11
- Distribution
- Worldwide Distribution -- USA, Sweden, Japan, Hong Kong, Italy, the Netherlands, Singapore, UK, and Spain.
- Recall number
- Z-1221-2010
- Recall status
- Terminated
- CDRH record ID
- 87748
- FDA classified
- 2010-05-06
- Recall event number
- 54165