Open original FDA record ↗Recall record
- FDA identifier
- Z-1221-2026
- Company in FDA record
- Beckman Coulter, Inc.
- FDA classification date
- Jan 30, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
Source facts
- Action
- During the week of 11/10/2025, the firm sent via postal and e-mail an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter has confirmed that specific diluent lot numbers of COULTER DxH Diluent and DxH ECO Diluent may contribute to elevated platelet (PLT) Daily Checks background counts cycle when used with the UniCel DxH 600/690T/800/900 analyzers. Customers are instructed to: "Check their COULTER DxH Diluent and DxH ECO Diluent inventory against the provided affected lot numbers. "If their lot numbers are NOT affected, continue routine operation and no further action is required. "If their lot numbers are affected, perform a Daily Checks cycle. o Each container must be individually verified through Daily Checks before use according to the instructions below. o If Daily Checks pass, continue routine operation and no further action is required, on the active diluent container(s), currently in use. "If the Daily Checks cycle does not pass, follow the instruction provide in the Recall Letter. For questions or further assistance - email BECFA@beckman.com
- Initiated
- 2025-11-10
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Canada, Chile, Mexico, Saint Kitts and Nevis, Trinidad and Tobago, and Viet Nam.
- Recall number
- Z-1221-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217394
- FDA classified
- 2026-01-30
- Recall event number
- 97992