Open original FDA record ↗Recall record
- FDA identifier
- Z-1222-2012
- Company in FDA record
- Biomet U.K., Ltd.
- FDA classification date
- Mar 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This recall was initiated following a report that a bag labeled as REF. 32-487557/Lot no 2561334 (Vanguard Tibial Resection head universal w/ alignment tower) actually contained a Ref. 32-487002/Lot no. 2561329 (Vanguard Distal cutting Block w/out handle attachment). Investigation found packaging of the affected product may contain the incorrect item.
Source facts
- Action
- Biomet U.K. sent a Field Safety Corrective Action notice dated February 14, 2012 to all affected customers. The notice identified the affected products, problem and actions to be taken. Customers were instructed to locate, discontinue use and quarantine the recalled products pending return to Biomet. The letter recommended customers provide the notice to any other organization who may have received the affected instruments. Customers were instructed to complete and return without delay the attached "Fax-Back Response Form" acknowledging receipt of notice and indicating the quantity of products to be returned. For questions call +44 (0) 1656 761658, fax: +44 (0) 1656 645454 and e-mail uk.complaints@biomet.com
- Initiated
- 2012-02-14
- Root cause
- Packaging change control
- Terminated
- 2012-11-06
- Distribution
- Worldwide Distribution - in the countries of France, Japan, Italy, Netherlands, UK , no devices distributed in the USA
- Recall number
- Z-1222-2012
- Recall status
- Terminated
- CDRH record ID
- 107699
- FDA classified
- 2012-03-19
- Recall event number
- 61274