Skip to content

Public FDA record

Z-1222-2018: QUANTA-Lyser 240 EIA, Part #GS0241. The instrument is a fully automated, high-throughput, bench top syst

Recall record Z-1222-2018 for Inova Diagnostics Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1222-2018
Company in FDA record
Inova Diagnostics Incorporated
FDA record creation date
Mar 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device power supply was not manufactured according to specification which can result in the loss of the protective grounding of the instrument.

Source facts

Action
The letters were issued via email on January 26, 2018, informing the consignee of the issue and the actions to be taken. The response forms were requested to be completed and returned by February 23, 2018, indicating receipt of the notification and confirmation that the power supply has been replaced. The firm reported on February 23, 2018 that responses have been received from both consignees. For further questions, please call (858) 586-9900 x 1381.
Initiated
2018-01-26
Root cause
Nonconforming Material/Component
Terminated
2018-05-22
Distribution
There was no U.S. distribution. Distribution was made to Panama and United Kingdom.
Recall number
Z-1222-2018
Recall status
Terminated
CDRH record ID
162184
FDA record created
2018-03-27
Recall event number
79380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗