Z-1222-2022: DYNEX Agility, Agility Analyzer, Model No. 67000

Recall record Z-1222-2022 for Dynex Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1222-2022
Company in FDA record
Dynex Technologies, Inc.
FDA record creation date
Jun 2, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the correct control, which may lead to delayed patient results.

Source facts

Action
The firm first notified customers via mail and email between June and July 2020. Although only 17 units had been released with v1.4.3 installed, the firm notified all customers in possession of a device out of an abundance of caution. Customers were given workaround instructions to prevent the problem from occurring while the software was being corrected. Customers then received a follow up Field Safety Notice dated September 29, 2020. Distributors were asked to pass on the notification to any downstream customer accounts. The Agility software was updated to v1.4.7 to prevent controls and calibrators from sharing the same name. The update should prevent recurrence of the issue.
Initiated
2020-07-05
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, DC, FL, GA, IL, IN, MA, MD, MI, NE, NJ, NY, OH, PA, PR, TN, TX, UT, VA, WI, WV and the countries of Belgium, Brazil, China, Czech Republic, France, Germany, Mexico, Qatar, Romania, Russia, South Korea, Switzerland, Taiwan, and UK.
Recall number
Z-1222-2022
Recall status
Open, Classified
CDRH record ID
193396
FDA record created
2022-06-02
Recall event number
90099

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗