Open original FDA record ↗Recall record
- FDA identifier
- Z-1222-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Jan 30, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
Source facts
- Action
- On December 17, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Carefully read the content of this notification. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the instruction for intermittent output activation and applicable Warnings and Cautions. 3. Keep a copy of this notification with the Instructions for Use for any affected devices remaining in your inventory. 4. If you have further distributed the affected product, please identify your customers and forward this notification to them. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
- Initiated
- 2024-02-17
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1222-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217415
- FDA classified
- 2026-01-30
- Recall event number
- 98007