Open original FDA record ↗Recall record
- FDA identifier
- Z-1223-2011
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Feb 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
Source facts
- Action
- Product bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations.
- Initiated
- 2005-10-01
- Root cause
- Other
- Terminated
- 2011-03-02
- Distribution
- Worldwide Distribution
- Recall number
- Z-1223-2011
- Recall status
- Terminated
- CDRH record ID
- 96778
- FDA classified
- 2011-02-14
- Recall event number
- 57607