- FDA identifier
- Z-1224-03
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Sep 11, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator. GLU slides are processed by high volume chemistry systems typically used in hospital laboratories.
Source facts
- Action
- GLU Slides: Letters dated between 7/28/03 and 8/11/03 provide revised storage instructions. Customers using NaF/KOx blood collection tubes and have VITROS GLU slides that have been refrigerated for more than four months will receive replacement product. Those customers are instructed to continue to use the slides only until replacement product arrives. GLU DT Slides: Letters dated between 7/28/03 and 8/11/03 with the same instructions as for GLU slides.
- Initiated
- 2003-07-28
- Root cause
- Other
- Terminated
- 2003-11-24
- Distribution
- Nationwide and worldwide distribution to distributors and end-users.
- Recall number
- Z-1224-03
- Recall status
- Terminated
- CDRH record ID
- 29040
- FDA classified
- 2003-09-11
- Recall event number
- 27035