A complaint was received for Item 42-5320-067-02, lot 63236047. It was report that there was no set screw or taper plug included in the box. Complaint investigation confirmed that the entire lot of 63236047 was missing the set screw and taper plug.
Source facts
Action
On December 12, 2017 an URGENT MEDICAL DEVICE RECALL LETTER was issued to customers will the following guidance:
1. Review this notification and ensure that affected personnel are aware of the contents.
2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.
Questions or concerns can be directed to 1-574-371-3071.
Initiated
2017-12-12
Root cause
Under Investigation by firm
Terminated
2019-04-09
Distribution
Non-USA Customer
Recall number
Z-1224-2018
Recall status
Terminated
CDRH record ID
162203
FDA record created
2018-03-27
Recall event number
79394
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.