Skip to content

Public FDA record

Z-1224-2024: VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)

Recall record Z-1224-2024 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1224-2024
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Mar 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Source facts

Action
On January 26, 2024, the firm sent a customer letter to all customers who have V3.8.1 installed on their VITROS Analyzer. Customers were informed of this issue and instructed to follow the instructions when using the VITROS System onboard QC package to update the QC baseline. The software will be updated to mitigate both issues within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Initiated
2024-01-26
Root cause
Software Design Change
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Recall number
Z-1224-2024
Recall status
Open, Classified
CDRH record ID
205912
FDA classified
2024-03-01
Recall event number
93925

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗