Open original FDA record ↗Recall record
- FDA identifier
- Z-1225-2022
- Company in FDA record
- MicroPort Orthopedics Inc.
- FDA record creation date
- Jun 2, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.
Source facts
- Action
- The firm sent out a recall notification to the distributors on 05/04/2022 via email. Phone calls were also made to the distributors. Additionally, the US field communication letter was mailed to the surgeon and facility in regard to the one case of implantation on May 10, 2022. The consignee letter informed the consignee to cease use and distribution of the affected product, check inventory for affected product, complete the response form, return affected product and notify affected consignees. The field surgeon letter and hospital letter informs the consignee to cease use and distribution of the affected product, monitor patients per normal protocol, return affected product and sign and return the response form.
- Initiated
- 2022-05-04
- Root cause
- Process change control
- Terminated
- 2024-01-16
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of FL, IN, MO, CA and the country of Canada.
- Recall number
- Z-1225-2022
- Recall status
- Terminated
- CDRH record ID
- 193557
- FDA record created
- 2022-06-02
- Recall event number
- 90140