Z-1225-2023: CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provide

Recall record Z-1225-2023 for St. Jude Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1225-2023
Company in FDA record
St. Jude Medical
FDA classification date
Mar 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

Source facts

Action
Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to do the following: 1. Prior to the implantation procedure, sensor data should be entered, and the clinician should proceed to the next screen to complete the Error #8 screening, prior to venipuncture. 2. If an Error #8 message occurs, the sensor should not be implanted. 3. Work with your Abbott Sales Representative to exchange the affected sensor. 4. Review Appendix B for labeling and supplemental information related to implantation and monitoring of CardioMEMS PA Sensors.
Initiated
2023-02-07
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Recall number
Z-1225-2023
Recall status
Open, Classified
CDRH record ID
198325
FDA classified
2023-03-06
Recall event number
91606

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗