Skip to content

Public FDA record

Z-1226-2019: Concorde Lift, Lordotic Expandable Interbody Device, spinal fusion implant 9x23 (197809023L), 9x27 (19780

Recall record Z-1226-2019 for DePuy Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1226-2019
Company in FDA record
DePuy Spine, Inc.
FDA classification date
Apr 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A complaint trend was observed for post-operative loss of height and migration for Concorde Lift G3+ implants.

Source facts

Action
On 4/8/19, the firm distributed Medical Device Recall letters to its customers. The letter recommended that surgeons use alternative implants. In addition, customers were asked to immediately case use of the affected devices and to return all affected implants on hand. In addition, customers were asked to complete and return the business reply form. Customers were asked to notify surgeon users and others in the facility who need to be informed of the recall. If any affected product was forwarded to another facility, please immediately contact that facility to communicate the field action. For product-related questions, please contact your local DePuy Synthes Spine Sales Consultant. For questions about device recall information provided, please contact Stephanie Koenig, Recall Associate, at 574 371-4584 (M-F; 8 a.m. 5 p.m. EST).
Initiated
2019-04-08
Root cause
Under Investigation by firm
Terminated
2023-01-12
Distribution
US Nationwide distribution. Foreign distribution to Austria, Switzerland, and United Kingdom.
Recall number
Z-1226-2019
Recall status
Terminated
CDRH record ID
171927
FDA classified
2019-04-27
Recall event number
82587

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗