Z-1226-2020: VITROS Chemistry Products Calibrator Kit 2- IVD calibrator for VITROS 250/350/5,1 FS/4600/XT 3400 Chemis

Recall record Z-1226-2020 for Ortho Clinical Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1226-2020
Company in FDA record
Ortho Clinical Diagnostics Inc
FDA record creation date
Feb 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for Negatively Biased Na+ Urine Results Using Several Lots of VITROS Chemistry Products Calibrator Kit 2

Source facts

Action
Ortho Clinical Diagnostics (Ortho) 16 December 2019, customer letter (Ref. CL2019-343) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail.Letter states reason for recall , health risk and provided with required actions depending upon whether VITROS Calibrator Kit 32 was available for use. VITROS Calibrator Kit 32 is available for use in the U.S. therefore the required actions in order to process urine samples on VITROS Na+ Slides are to:calibrate. o discontinue using the calibration for urine sodium (only) obtained using an affected VITROS Calibrator Kit 2 lot until either VITROS Na+ Slide GEN 27 is available and/or GEN 26 performance is updated in mid-late January 2020 (refer to Resolution section). Ortho will expedite 1 sales unit of VITROS Calibrator Kit 32, if needed. Request via Confirmation of Receipt form. o Use an alternate urine sodium method until calibration with VITROS Calibrator Kit 32 is complete. o If no alternate urine sodium method is available for use, consult with your medical director to determine if the urine sodium calibration obtained using an affected VITROS Calibrator Kit 2 lot is acceptable for continued use until: Calibration with VITROS Calibrator Kit 32 is complete or VITROS Na+ Slides GEN 26 performance is updated (refer to Resolution section of the communication). VITROS Calibrator Kit 2 is acceptable to use after this occurs or VITROS Na+ Slides GEN 27 or higher is available. VITROS Calibrator Kit 2 will be acceptable to use. Questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Initiated
2019-12-16
Root cause
Under Investigation by firm
Terminated
2020-11-12
Distribution
Nationwide Foreign: Australia Brazil Canada L3R 4G5 Chile China Colombia India Japan Mexico Singapore United Kingdom France Germany Italy Spain Portugal Poland Russia Denmark Norway Sweden Belgium Netherlands
Recall number
Z-1226-2020
Recall status
Terminated
CDRH record ID
178924
FDA record created
2020-02-13
Recall event number
84654

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗