Skip to content

Public FDA record

Z-1228-2015: CareLink Pro Medtronic CareLink Pro MMT-7335 is a personal computer software application designed to enha

Recall record Z-1228-2015 for Medtronic MiniMed Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1228-2015
Company in FDA record
Medtronic MiniMed Inc.
FDA classification date
Mar 2, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The firm is informing customers of a software error that might result in an inaccurate display of the Temp Basal time and duration in CareLink Clinical reports with data uploaded from the MiniMed 620G or 640G insulin pump.

Source facts

Action
Medtronic MiniMed Inc. sent an Urgent Medical Device Correction letter in December 2014 to all affected customers. The firm states in their letter that patients can continue to use their MiniMed 620G and 640G insulin pump systems, and that there is no risk to patients caused by this issue. The firm instructs that when reviewing the CareLink Pro 4.0 or CareLink Clinicial reports from the MiniMed 620G and 640G insulin pumps not to use the Temp Basal graphic for clinical consideration. For questions regarding this recall call 818-567-4700.
Initiated
2014-11-10
Root cause
Software design
Terminated
2015-03-10
Distribution
Worldwide Distribution to Japan only. CareLink Clinical is for clinical trials only (US and International)
Recall number
Z-1228-2015
Recall status
Terminated
CDRH record ID
133579
FDA classified
2015-03-02
Recall event number
70424

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗