Z-1228-2018: ACE Trochanteric Nail System STER TROC LAG SCR 70 MM The Trochanteric Nail System is intended to treat s

Recall record Z-1228-2018 for Zimmer Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1228-2018
Company in FDA record
Zimmer Biomet, Inc.
FDA classification date
Mar 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

Source facts

Action
On December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071.
Initiated
2017-12-20
Root cause
Device Design
Terminated
2020-04-22
Distribution
USA (nationwide)
Recall number
Z-1228-2018
Recall status
Terminated
CDRH record ID
162243
FDA classified
2018-03-14
Recall event number
79408

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗