Z-1228-2023: Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT

Recall record Z-1228-2023 for Covidien, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1228-2023
Company in FDA record
Covidien, LP
FDA classification date
Mar 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.

Source facts

Action
On January 25, 2023, the firm notified customers via Urgent Medical Device Recall letter. Customers were asked to take the following immediate actions: 1. Identify and quarantined all unused affected product. 2. Return unused affected product to Medtronic using the Customer Confirmation Form. 3. Forward the information to all who need to be aware within your organization, or to any organization where the affected product was transferred. There are no additional actions required for patients where a stapler in scope of this recall was used during a procedure. These patients should continue to be monitored in accordance with your medical facility s standard care protocols. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.
Initiated
2023-01-25
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide distribution in the states of Arkansas, California, Florida, Georgia, Louisiana, Massachusetts, New York, North Carolina, Ohio, Washington and the countries of Australia, Austria, Belgium, Canada, Canary Islands, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Caledonia, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Recall number
Z-1228-2023
Recall status
Open, Classified
CDRH record ID
198417
FDA classified
2023-03-06
Recall event number
91650

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗