Open original FDA record ↗Recall record
- FDA identifier
- Z-1228-2025
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Feb 21, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
Source facts
- Action
- On January 8, 2025, CareFusion 303 issued a "Urgent: Medical Device Correction" Notification to affected customer via mail or Email. CareFusion asked consignees to take the following actions: 1. Customers should immediately ensure that MedBank" devices are not installed in the procedural or operating space and at least 1.5 meters from patients. 2. Report any issues associated with the above products to BD Pyxis" MedBank" Product Support Line via phone at 866-930-9251. 3. Report any adverse health consequences experienced with the use of this product to the FDA's MedWatch Adverse Event Reporting program. 4. Please post these notices with the device until labeling has been updated. 5. Complete the attached Customer Response Form and return to BD.
- Initiated
- 2025-01-08
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
- Recall number
- Z-1228-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212271
- FDA classified
- 2025-02-21
- Recall event number
- 96124