Metal guidewires used in Hemodialysis Catheters may have been reverse loaded within the dispenser. Potential exists for the stiff end of the guidewire to be inserted into the patient instead of the flexible end.
Source facts
Action
Consignees were notified by letter on 08/10/2007 and asked to remove and return all unused affected product to Bard.
Initiated
2007-08-10
Root cause
Other
Terminated
2008-04-25
Distribution
Worldwide, including USA, Canada, Latin America, and Europe (Austria, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, and Sweden).
Recall number
Z-1229-2007
Recall status
Terminated
CDRH record ID
54054
FDA classified
2007-09-20
Recall event number
38621