Skip to content

Public FDA record

Z-1229-2011: AMS brand GreenLight PV Laser System, Model-P/N: 0010-9230; Product is manufactured and distributed by AM

Recall record Z-1229-2011 for AMS Innovative Center - San Jose. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1229-2011
Company in FDA record
AMS Innovative Center - San Jose
FDA classification date
Feb 14, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A failure mode whereby the fiber attachment to the GreenLight PV laser system could inadvertently be broken. If fiber were to bend sharply at point of fiber connection, it may break and cause a burn.

Source facts

Action
AMS Innovative Center - San Jose issued Important Product Update Information letters via express mail on February 8, 2010, and September 15, 2010 to all its direct consignees. The letters identified the product, the problem, and the action to be taken by the customer. Customers were instructed to attach the Fiber Guard to their GreenLight PV Laser System, and to complete the Response Form in the enclosed, self-addressed, postage paid envelope to AMS. For any questions regarding this recall call the Customer Care Department at 1-800-356-7600 or send an e-mail to CustomerCare@ammd.com.
Initiated
2010-02-08
Root cause
Other
Terminated
2011-02-14
Distribution
Worldwide Distribution including US
Recall number
Z-1229-2011
Recall status
Terminated
CDRH record ID
92557
FDA classified
2011-02-14
Recall event number
56058

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗