The monitor may not alarm appropriately for a pediatric or neonatal patient.
Source facts
Action
Philips sent each known affected customer an Urgent Medical Device Correction Notification/Field Safety Notice dated December 2016 to inform customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue.
The correction will consist of providing a replacement MRx unit, free of charge, for all units affected by this issue. When the replacement unit arrives, please use the return shipping label and packaging provided with the replacement MRx unit to return the affected unit to Philips.
Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice:
Option A:
For any neonatal or pediatric patient, if the bedside monitor (without CO2) is connected to CMS200, do not place the bedside monitor in Standby when the patient is transported to a procedure. This will allow the alarms limits to work properly. Contact your service provider to determine the settings of your system and change if necessary.
Option B:
For any neonatal or pediatric patient, if the bedside monitor (without CO2) connected to CMS200 is placed in Standby when the patient is transported to a procedure. When the same patient returns to unit and is placed back on monitor, take the following steps:
If clinician need disable or change alarm limit at CMS200 side, he/she need to discharge this patient at CMS200 side first, and admit same patient ID again. (Reference: Instruction for Use, Document Number 4535 645 61281, Page 89 of 199, section: Discharging a patient) This will allow the alarms limits to work properly.
Contact your service provider to determine the settings of your system and change if necessary.
For questions regarding this recall call 978-659-3000.
Initiated
2017-01-17
Root cause
Unknown/Undetermined by firm
Terminated
2019-10-10
Distribution
Worldwide distribution including
Austria
Bangladesh
Bulgaria
Chile
France
Germany
Hong Kong
India
Ireland
Israel
Italy
Korea, Republic of
Malaysia
Myanmar
Philippines
Poland
Portugal
Saudi Arabia
Thailand
Turkey
United Arab Emirates
United Kingdom
United Republic of Tanzania
Recall number
Z-1229-2017
Recall status
Terminated
CDRH record ID
152614
FDA record created
2017-02-22
Recall event number
76274