Z-1229-2023: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-

Recall record Z-1229-2023 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1229-2023
Company in FDA record
Datascope Corp.
FDA classification date
Mar 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.

Source facts

Action
URGENT MEDICAL DEVICE CORRECTION dated 2/7/23 were sent to customers. There have been reported failures of the high pressure helium regulator which may cause a helium leak in the Cardiosave Hospital Cart. The high pressure helium regulator is located in the Cardiosave Hospital Cart and regulates the helium pressure of the external helium supply. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir. Please refer to images under Issue 3 for reference. Issue 3: Leak in the High Pressure Helium Regulator Identification of the issue: There have been reported failures of the high pressure helium regulator which may cause a helium leak in the Cardiosave Hospital Cart. The high pressure helium regulator is located in the Cardiosave Hospital Cart and regulates the helium pressure of the external helium supply. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir. Please refer to images below for reference. Datascope/Getinge has received 51 reported complaints of helium leak as a result of high pressure helium regulator failure over a 2 year period. There have been 0 adverse events reported. Risk to Health: Should a Cardiosave s helium supply be depleted due to an impaired helium pressure regulator, therapy may be interrupted. As with any therapy interruption, the degree of subsequent hemodynamic stability is related to the patient s overall clinical condition, those critically ill are more vulnerable to clinical decline. The risk of therapy interruption from a depleted helium supply is mitigated by the advanced notice provided to the User (a minimum of approximately 24 hours). The Pump Console s internal helium
Initiated
2023-02-07
Root cause
Under Investigation by firm
Distribution
US Nationwide. Global Distribution.
Recall number
Z-1229-2023
Recall status
Open, Classified
CDRH record ID
198400
FDA classified
2023-03-06
Recall event number
91645

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗