Open original FDA record ↗Recall record
- FDA identifier
- Z-1230-2012
- Company in FDA record
- Biosense Webster, Inc.
- FDA classification date
- Mar 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.
Source facts
- Action
- Three Urgent Customer Notification recall communications were initiated 2/9/12 with Biosense forwarding the notification to customers with affected backplane cards (electrophysiology lab managers, a notification for customers with affected back plane cards (risk managers), and a notification for customers with affected backplane cards and smarttouch module along with customer acknowledgement forms. The letters provides the customers with an explanation of the problem identified and steps to take in the event that the issue is observed. If the steps provided do not resolve the issue, customers are to contact their Biosense Webster representative. A Biosense Webster representative will contact customers to schedule an appointment for the replacement of the Backplane card. Customers were instructed to complete and return the attached Acknowledgement Form in accordance with the instructions listed on the form. If customers have questions, they can contact their BWI representative.
- Initiated
- 2012-02-09
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-09-06
- Distribution
- Worldwide Distribution -- US, Japan, China, Ireland, Germany, Italy, Slovenia, Indonesia, Poland, Spain, United Kingdom, Switzerland, Belgium, and Canada.
- Recall number
- Z-1230-2012
- Recall status
- Terminated
- CDRH record ID
- 107549
- FDA classified
- 2012-03-20
- Recall event number
- 61197