There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
Source facts
Action
Siemens Healthcare Diagnostics initiated a customer communication for this issue via letter beginning December 18, 2019.
Initiated
2019-12-18
Root cause
Under Investigation by firm
Terminated
2020-06-25
Distribution
US Nationwide
Recall number
Z-1230-2020
Recall status
Terminated
CDRH record ID
178848
FDA classified
2020-01-24
Recall event number
84612