Skip to content

Public FDA record

Z-1230-2025: 139088-01 BD PYXIS MEDBANK MEDPASS MODULE 169-151 BD PYXIS MEDBANK MINI 1FH-2HH MEDPASS 169-152 BD PYXIS

Recall record Z-1230-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1230-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Feb 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Source facts

Action
On January 8, 2025, CareFusion 303 issued a "Urgent: Medical Device Correction" Notification to affected customer via mail or Email. CareFusion asked consignees to take the following actions: 1. Customers should immediately ensure that MedBank" devices are not installed in the procedural or operating space and at least 1.5 meters from patients. 2. Report any issues associated with the above products to BD Pyxis" MedBank" Product Support Line via phone at 866-930-9251. 3. Report any adverse health consequences experienced with the use of this product to the FDA's MedWatch Adverse Event Reporting program. 4. Please post these notices with the device until labeling has been updated. 5. Complete the attached Customer Response Form and return to BD.
Initiated
2025-01-08
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Recall number
Z-1230-2025
Recall status
Open, Classified
CDRH record ID
212273
FDA classified
2025-02-21
Recall event number
96124

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗