Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it
is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.
Source facts
Action
A recall notification is being forwarded to all users and consignees who have received product. The notification of this voluntary recall will be sent to the following consignee in each facility:+
Director of Operating Room Services
Proposed Level of Effectiveness Checks:
Terumo CVS will verify that 100% of all letters are delivered to affected consignees
Initiated
2015-01-05
Root cause
Other
Terminated
2017-02-23
Distribution
The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.
Recall number
Z-1232-2017
Recall status
Terminated
CDRH record ID
152937
FDA classified
2017-02-22
Recall event number
76346