Z-1232-2025: LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.

Recall record Z-1232-2025 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1232-2025
Company in FDA record
Physio-Control, Inc.
FDA classification date
Feb 25, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect keypad was incorrectly installed onto a defibrillator/monitor.

Source facts

Action
On 02/03/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via mail to a customer inform them that Stryker discovered that a keypad was incorrectly installed onto a LIFEPAK 20e defibrillator/monitor. There is a possibility that the incorrectly installed keypad may cause difficulty or the inability to actuate the keypad or buttons resulting in the inability to utilize the device. Additionally, there is a chance of water ingress or contamination resulting in damage to the internal machinery of the device. Customers are instructed to: 1. Immediately check your internal inventory to locate the product listed on the attached business reply form. 2. Please remove and quarantine the affected product from use. 3. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, please email the enclosed Business Reply Form (BRF) by January 31st, 2025 date, to RSRecall@Stryker.com. 4. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return and replacement of your product(s). 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until replacement. 6. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: rsdevicetracking@stryker.com b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ For question or concerns - contact Customer Service +1 800 442 1142 or by email RSRecall@Stryker.com.
Initiated
2025-02-03
Root cause
Employee error
Distribution
U.S.: WV O.U.S.: N/A
Recall number
Z-1232-2025
Recall status
Open, Classified
CDRH record ID
211995
FDA classified
2025-02-25
Recall event number
96095

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗