Z-1232-2026: Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System Model/Ca

Recall record Z-1232-2026 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1232-2026
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Feb 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction

Source facts

Action
On 11/07/2025 and 11/13/2025, the firm emailed "Urgent Stop Use: Product Safety Notice" letters to customers informing them to stop use of specific da Vinci Surgical Systems. In 2022 Intuitive Surgical initiated a field action related to instrument arm ball screw failures. As a result, a field test was implemented to improve detection of affected ball screw. Intuitive has become aware that this test did not alert the service technicians of pass/fail results. Customer are instructed to immediately stop using affected da Vinci Systems. Intuitive will provide a formal follow-up communication, and our representative will contact you to schedule a site visit to perform the necessary corrections. Please distribute this communication to all relevant hospital personnel (e.g., surgeons, OR directors, risk managers, etc.) who use your affected da Vinci Surgical System. For Questions, contact Customer Service at: " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Initiated
2025-11-07
Root cause
Process control
Distribution
U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, Netherlands, South Korea, Spain, Switzerland, Taiwan and United Kingdom
Recall number
Z-1232-2026
Recall status
Open, Classified
CDRH record ID
217501
FDA classified
2026-02-02
Recall event number
98050

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗