Open original FDA record ↗Recall record
- FDA identifier
- Z-1233-2008
- Company in FDA record
- Arjo, Inc.
- FDA classification date
- Jul 18, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The ceiling lift may not stop lateral movement after releasing the left (or right) action button on a ceiling lift equipped with an infrared hand control.
Source facts
- Action
- Arjo sent Urgent Device Field Correction Notification letters dated 2/15/08 on the BHM Medical letterhead to the end user accounts who received the affected IR remotes, advising them of the potential for the remote to become difficult to operate or to create uncontrolled movement of the ceiling lift. The accounts were instructed to remove the IR remote from service until the field correction is issued by BHM Medical. Instructions for removal of the IR equipment were provided in the Safety Advisory Notice included with the letter. The accounts were requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter, and indicating the number of units requiring correction.
- Initiated
- 2008-02-20
- Root cause
- Device Design
- Terminated
- 2010-09-11
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1233-2008
- Recall status
- Terminated
- CDRH record ID
- 68292
- FDA classified
- 2008-07-18
- Recall event number
- 46800