Skip to content

Public FDA record

Z-1233-2015: ARROW FiberOptix Intra-Aortic Balloon Catheter Kit, 40 cc 7.5 Fr Product Code: IAB-05840-LWS The Arrow IA

Recall record Z-1233-2015 for Arrow International Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1233-2015
Company in FDA record
Arrow International Inc
FDA classification date
Mar 2, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Outer package product label incorrectly identifies the product as FiberOptix Flex IAB 7.5Fr rather than the correct FiberOptix IAB 8Fr

Source facts

Action
Arrow International notified Users, Distributors, Risk Managers by letter dated February 5, 2015. The letter explains the labeling error and requests immediately discontinue use and quarantine product. Customers were instructed to complete the Recall Acknowledgement Form and return by fax to 1-855-419-8507 whether they have or do not have the affected product. Customers with questions were instructed to contact their local sales representataive or Customers Service at 1-866-246-6990. For questions regarding this recall call 610-378-0131.
Initiated
2015-02-05
Root cause
Labeling mix-ups
Terminated
2016-10-26
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Belgium
Recall number
Z-1233-2015
Recall status
Terminated
CDRH record ID
133695
FDA classified
2015-03-02
Recall event number
70436

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗