Open original FDA record ↗Recall record
- FDA identifier
- Z-1233-2016
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA record creation date
- Mar 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.
Source facts
- Action
- Roche Diagnostics sent an "Urgent Medical Device Correction" letter via UPS Ground (receipt required) on February 23, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To prevent this malfunction, a software patch has been developed and validated. The installation of this software patch is mandatory and will be performed by a Roche Field Engineering Specialist (FES). Your Roche FES will contact you to schedule the installation of the software patch. Until the software patch is implemented, Roche recommends that instrument operators monitor their systems for the occurrence of Error 140 Gripper has lost tube. If this error occurs, all components in contact with a spill must have extra cleaning. Refer to the cobas p 512 pre-analytical instrument Operators Manual Version 1.5. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 for further questions.
- Initiated
- 2016-02-23
- Root cause
- Software design
- Terminated
- 2018-01-11
- Distribution
- US Distribution including Puerto Rico and to the states of :TX, OH, TN, AZ, WA, MI and GA
- Recall number
- Z-1233-2016
- Recall status
- Terminated
- CDRH record ID
- 143776
- FDA record created
- 2016-03-23
- Recall event number
- 73292