BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD.
FDA classification date
Feb 25, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.
Source facts
Action
On 1/10/2025, recall notices were mailed to customers who were asked to do the following:
1) Replace and discard any reagent bottles that were on-board the analyzer at the time of any software upgrade.
2) Review Quality Control data and actions taken, for shifts out of allowed laboratory range which could have indicated potential deterioration in reagent stability.
3) Firm recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted.
4) If specific software upgrade date is required, contact firm.
5) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter.
This issue will not occur with future software updates.
If you have any questions regarding this notice, please contact the firm's Customer Support Center:
Website: http://www.beckmancoulter.com
Hotline: (800) 854-3633
Initiated
2025-01-10
Root cause
Software change control
Distribution
US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR.
OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of Korea
Recall number
Z-1233-2025
Recall status
Open, Classified
CDRH record ID
212289
FDA classified
2025-02-25
Recall event number
96193
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.