Open original FDA record ↗Recall record
- FDA identifier
- Z-1233-2026
- Company in FDA record
- Encore Medical, LP
- FDA classification date
- Feb 2, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.
Source facts
- Action
- On 1/5/2026, recall notices were emailed to customers who were asked to do the following: 1) This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 2) Contact the firm to arrange for affected device return. 3. Complete and return the acknowledgement and response form via email to Lesli.Helmick@enovis.com Customers with questions can email the firm at productsafety@enovis.com
- Initiated
- 2026-01-05
- Root cause
- Labeling Change Control
- Distribution
- US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA
- Recall number
- Z-1233-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217793
- FDA classified
- 2026-02-02
- Recall event number
- 98177