Open original FDA record ↗Recall record
- FDA identifier
- Z-1234-2008
- Company in FDA record
- Arjo, Inc.
- FDA classification date
- Aug 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Repeated use of the ratchets that lock the position of the back and leg rest of the stretcher may damage the retention pin that prevents the latch from disengaging from the anchor system. This would allow the backrest to pivot freely down to the floor.
Source facts
- Action
- BHM Medical initially notified the end users on 1/23/07 of the potential for the back and leg rest latch ratchets to become damaged, allowing the rests to pivot freely. Should a patient be in the stretcher at the time, the patient could fall, potentially sustaining cuts, bruises or even more severe injuries. The letters provided safety instructions to follow until a permanent fix is available, and requested the accounts to complete and return the enclosed response form acknowledging receipt of the letter and dissemination of the safety instructions to all employees operating the lifts with the affected GK stretchers. BHM sent Customer Field Correction Notification letters dated 1/23/08 to the accounts, along with the Technical Advisory Notice - TAN A dated 1/3/08, reiterating the safety instructions from the 1/23/07 letter and requesting the accounts to, if at all possible, remove the equipment from service until an authorized engineer can replace the ratchets in the backrest and leg rest latches. All consignees have been contacted.
- Initiated
- 2007-01-23
- Root cause
- Device Design
- Terminated
- 2008-08-25
- Distribution
- Nationwide Distribution --- including states of California, Connecticut, Florida, Mississippi and Texas.
- Recall number
- Z-1234-2008
- Recall status
- Terminated
- CDRH record ID
- 67340
- FDA classified
- 2008-08-24
- Recall event number
- 46383