Open original FDA record ↗Recall record
- FDA identifier
- Z-1234-2009
- Company in FDA record
- Dentsply Worldwide Headqrt
- FDA classification date
- Apr 28, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product is manufactured with a left-hand thread instead of a right hand thread. This error could cause tooth breakage if the posts were removed without knowing that the threads were reversed.
Source facts
- Action
- Dentsply notified their distributors and dentists of the recall by email sent on 03/20/2009 by Urgent Recall Notice letter explaining the problem and requesting return of the product from their inventory and from their consignees. The letter requests faxback response (enclosed with the recall notification) within five (5) working days. The letter included a response page to ensure the receipt of the notice. A second follow up letter will be sent to all that do not respond to the first letter. The firm identified another distributor within the company, DENTSPLY Tulsa Specialty Products, who also sent out identical recall letters to those they distributed the product on 03/25/09.
- Initiated
- 2009-03-20
- Root cause
- Process control
- Terminated
- 2010-02-12
- Distribution
- Worldwide distribution: USA, Canada, Germany, France, Holland, Switzerland, The Netherlands, Italy, Kuwait, Poland, Argentina, Greece, Russia, Czech Republic, Belgium, Turkey, Hungary, India, and Colombia.
- Recall number
- Z-1234-2009
- Recall status
- Terminated
- CDRH record ID
- 80334
- FDA classified
- 2009-04-28
- Recall event number
- 51469