Z-1234-2022: STORZ KARL STORZ-ENDOSKOPE REF 11272VUE CMOS Video Cysto-Urethroscope REF 11272VUEK CMOS Video Cysto-Ure

Recall record Z-1234-2022 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1234-2022
Company in FDA record
Karl Storz Endoscopy
FDA record creation date
Jun 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The endoscope may experience a loss of image when activating monopolar electrocautery, due a manufacturing assembly process issue.

Source facts

Action
On April 29, 2022, KARL STORZ sent an "Urgent Medical Device Recall Notice" Letter via UPS and posting to their website "https://urldefense.com/v3/__https://www.karlstorzcustomerupdates.com/wp-content/uploads/2022/04/C-View-Recall-Customer-Notice.pdf__;!!Lw1uGqvHvtJ_psGG2Et-1voSwbo!bg5okg2Sm91SJG6RWtBKXJ258Z4YgfwOv4ZIuU_ZuW_tH2G1X_CNLBZeGB0YWAZEjDZ4j_-Hvlm6xzzWmhKnBU6U10FQtQHO$" informing customers that certain serial numbers of the Flexible Video Cysto-Urethroscope may exhibit temporary loss of image when monopolar electrocautery is activated. Customers who are in possession of the identified scopes should discontinue use of the affected scopes and return them to KARL STORZ for replacement. Customers can confirm that Endoscope is part of the Recall by comparing the serial number to the serial numbers list in the Appendix. In addition, if customers have already returned any affected scope(s) as part of the Reprocessing Methods Recall, they must still complete the enclosed Acknowledgment Form for this recall and return it as soon as possible via the contact information specified on the form. If customer's Acknowledgment Form for the Reprocessing Methods Recall indicated that they have access to an appropriate sterilization method to sterilize the affected flexible endoscopes, then they will be eligible to have their C-View Endoscope(s) replaced by KARL STORZ. If they have not returned their Acknowledgment Form for the Reprocessing Methods Recall or if their Acknowledgement Form indicated that they do not have access to an appropriate sterilization method, then KARL STORZ will not be able to replace their C-View Endoscope(s) at this time. For questions regarding this action, please use the contact information provided below for the recall execution partner, Sedgwick: Telephone: 877-877-0317 Fax: 877-597-9587 E-Mail: karlstorz2935@sedgwick.com
Initiated
2022-05-03
Root cause
Process control
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, VA, VT, WA, WI, WV, and WY. The country of Canada.
Recall number
Z-1234-2022
Recall status
Open, Classified
CDRH record ID
193173
FDA record created
2022-06-03
Recall event number
90025

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗