Open original FDA record ↗Recall record
- FDA identifier
- Z-1235-2008
- Company in FDA record
- American Medical Systems
- FDA classification date
- Jul 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Missing part: The units in two lots of In-Fast Ultra Transvaginal Bladder Neck Support System may contain only one bone screw, but should contain two as indicated on the labeling. Two screws are required to complete the procedure. The missing screw will result in the inability to complete the procedure as expected if there are no additional screws readily available.
Source facts
- Action
- All customers received phone calls on 1/28/08 explaining the error and associated risks and were followed up with detailed letters. The letter "AMS Important Product Recall" dated 1/28/08 described the product and problem. Customers were advised to return the enclosed "Recall Acknowledgement Form" to AMS along with any affected product that they have in inventory.
- Initiated
- 2008-01-28
- Root cause
- Process control
- Terminated
- 2010-10-13
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1235-2008
- Recall status
- Terminated
- CDRH record ID
- 68313
- FDA classified
- 2008-07-16
- Recall event number
- 46833